Ozone Analyzer Mini-Hicon

The Model Mini-HiCon is an innovative Ozone Analyzer that combines a state-of-the-art gas-phase ozone sensor with an advanced electronics control module in a small, rugged enclosure. It is ideal for measuring ozone process gas in a wide variety of applications with high (up to 400 g/m3) ozone concentrations.
The Model Mini-HiCon Analyzer incorporates IN USA’s innovative dual optical-path sample sensing technology (patent-pending) to provide unparalleled performance and virtually eliminate the need for a reference gas measurement.
The Model Mini-HiCon Analyzer provides powerful microprocessor-based control, display, and output capabilities. It includes an alphanumeric, backlit LCD; 0-10 VDC and 4-20 mA analog outputs (opto-isolated); bi-directional RS232 digital communications; remote AutoZero capability; and automatic compensation for sample pressure and temperature.
Features
- Pressure and temperature compensated
- Remote AutoZero • High accuracy UV Absorption method
- Dual optical-path sample sensing
- Minimal maintenance - no moving parts
- Semiconductor-grade wetted materials
- Compact enclosure
- Integrated sensor and microprocessor-based controller
- Panel mount
Technical Specification
| ANALYZER |
Model Mini-HiCon™ integrated sensor and controller;
microprocessor-based
|
Application
|
Ozone in gases; continuous |
Measuring Principle
|
UV Absorption; dual optical-path sample sensing* |
Pressure & Temperature
|
Automatic compensation for sample pressure and temperature |
UV Light Source
|
Low pressure mercury vapor lamp |
Units of Measure
|
g/m3, g/Nm3, % w/w |
Range
|
0 to 400 g/m3, 0 to 26% w/w (at NTP) |
Repeatability
|
0.5 g/m3 or 1% of reading, including zero drift;
(higher precision available)
|
Resolution
|
0.1 g/m3 or 0.01 % w/w |
Cycle Time
|
Continuous measurement; 0.5 sec or 2 Hz sampling rate |
Sample Flow
|
0.2 to 1 SLPM |
Sample Ports
|
1/4" SwagelokTM (others available) |
| Display |
20 character, alphanumeric, backlit LCD |
Analog Outputs
|
4 to 20 mA and 0 to10 VDC, opto-isolated |
Digital I/O
|
RS-232-C, bi-directional |
Diagnostics
|
Internal diagnostics; instrument error relay |
AutoZero
|
Remote AutoZero Standard |
| Power |
90 to 240 VAC 50/60 Hz |
Dimensions
|
5.6” W x 2.8” H x 9.5” D
(143 x 71 x 241 mm)
|
| Maintenance |
UV lamp replacement every 24 months.
No other scheduled maintenance.
|
Compliance
|
CE |
|
|
|
|
|
|
|
|